12 ErrorMessages
-399 to -300: Device-Specific Errors ESG Family Signal Generators
Out of memory; There is not enough memory to create the
modulation. Baseband cannot function.
There is not enough memory to create the modulation map. Reduce the
size of any list/sweep, and preset the source to clear up any
inconsistencies. If the problem persists report this error to the nearest
Agilent Technologies sales or service office.
Out of memory; Object Memory Area
Indicates that memory was exhausted during instrument power-on.
Report this error to the nearest Agilent Technologies sales or service office.
Out of memory; Pattern exceeds digital modulation memory size.
If a user file is selected, it may be too large.
The user has specified a data file or other generation parameters that
would require more data generation pattern memory than is available
with the particular data generator installed in the instrument.
Out of memory; PRBS xx/xx
There was not enough memory to apply a scramble to data for the
baseband generator. In this case, the scramble is not applied to the
generator.
Out of memory; Shape subsystem unable to operate.
There is insufficient working memory available. Report this error to the
nearest Agilent Technologies sales or service office.
Out of memory; The table editor cannot function properly until
more memory is available.
There is insufficient working memory available. Reduce the size of any
sweep lists and try again.
Out of memory; The table editor cannot function properly until
more memory is available.
There is insufficient memory for the ARB to generate the requested
waveform.
Out of memory; Unable to allocate memory for SCPI response.
There is not enough memory to execute this SCPI query. Reduce the size of
any list/sweep, and preset the source to clear up any inconsistencies. If the
problem persists report this error to the nearest Agilent Technologies sales
or service office.