56 ErrorMessages
201 to 702: Device-Specific Errors ESG Family Signal Generators
A STATe: file could not be saved due to insufficient space, file corruption,
or another related problem.
I/O error; Trailing zero found in <filename>. Fixing...
A compressed state file has a zero at its end. This is a sign of file
corruption. The device fixes the problem by concealing the zero such that it
no longer triggers an error message. The file may be corrupt or unusable.
I/O error; The user file is too large.
The user file selected for transmission as the DCH data must be less than
or equal to 4096 bytes. Select a smaller user file.
I/O error; The user file is too large.
The user file selected for transmission as the TPC pattern must be less
than or equal to 256 bytes. Select a smaller user file.
I/O error; Unable to delete saved state from non-volatile
memory. No instrument state change.
A STATe: file could not be deleted due to the file not being found, file
corruption, or another file-related problem. If the file is displayed by a
memory catalog, delete it explicitly.
I/O error; Unable to read bit file.
There is an error reading the specified bit file. Try selecting the file again.
I/O error; Unable to recall from non-volatile memory. No
instrument state change.
The state file is not readable and the recall was aborted.
214 Not owner; Unable to delete saved state from non-volatile
memory. No instrument state change.
The user has attempted to write to a read-only memory subsystem.
218 I/O error; filename
An invalid file name was detected during ARB segment or sequence
processing.
501 Attenuator hold setting over range; Frequency change forced
attenuator adjust.