Agilent Technologies E4400-90333 Portable Generator User Manual


 
62 ErrorMessages
201 to 702: Device-Specific Errors ESG Family Signal Generators
DSP times out; Demodulator communication failed
The digital signal processor (DSP) on the Demodulator board (Option 300)
failed to respond within the appropriate amount of time. If cycling power
does not correct the problem, report this error to the nearest
Agilent Technologies sales or service office.
DSP times out; FlexDG communication failed
The instrument is unable to communicate with the FLEXDG data
generator board. The baseband generator can not function.
DSP times out; FPGA download failed: DUAL ARB BOARD
The initialization of hardware necessary for the operation of the dual
arbitrary waveform generator (Option UND) failed. The dual arbitrary
waveform generator is not functional. Report this error to the nearest
Agilent Technologies sales or service office.
DSP times out; Option 300 BERT capability
The digital signal processor (DSP) on the bit error rate tester board failed
to respond while Option 300 is operating. Report this error to the nearest
Agilent Technologies sales or service office.
607 DSP returns error.
The digital signal processor (DSP) is in an indeterminate state. Report this
error to the nearest Agilent Technologies sales or service office.
DSP returns error; ARB address line error (code)
The dual arbitrary waveform generator (Option UND) digital signal
processor (DSP) detected an internal error when processing a command.
The DSP specific error code is displayed in the parenthesis. Report this
error to the nearest Agilent Technologies sales or service office.
DSP returns error; ARB Command not recognized (code)
The dual arbitrary waveform generator (Option UND) digital signal
processor (DSP) detected a communications error with the command sent
to it. The DSP specific error code is displayed in the parenthesis. Report
this error to the nearest Agilent Technologies sales or service office.
DSP returns error; ARB communications failed. (code)
The dual arbitrary waveform generator (Option UND) digital signal
processor (DSP) detected an error and it is in an indeterminate state. The
DSP specific error code is displayed in the parenthesis. Report this error to
the nearest Agilent Technologies sales or service office.