Agilent Technologies E4400-90333 Portable Generator User Manual


 
Error Messages 39
ESG Family Signal Generators -299 to -200: Execution Errors
Settings conflict; Incompatible step sweep frequency setting.
Start, stop, center or span setting changed to be compatible
with last command.
Incompatible step sweep frequency setting. Start, stop, center or span
setting changed to be compatible with last command.
Settings conflict; Incompatible step sweep power setting.
Start, stop, center or span setting changed to be compatible
with last command.
Incompatible step sweep power setting. Start, stop, center or span setting
changed to be compatible with last command.
Settings conflict; Internal burst cannot be on with pulse
modulation.
There is a hardware conflict between internal burst and pulse modulation.
The most recently requested modulation will be turned on, the previous
modulation will be turned off.
Settings conflict; Invalid Channel Code. Channel code has been
changed to the highest channel code for this symbol rate.
An attempt was made to input a channelization code beyond the
limitations imposed by the symbol rate. The channelization code is reset to
the highest allowable code for the present symbol rate setting.
Settings conflict; Invalid Channel Code on channel <NUMBER>.
The channel code has been set to <NUMBER>.
The channelization code range has been changed due to new slot format
constraints. The channelization code has been set to the new upper limit.
The new slot format may affect all DPCH channels.
Settings conflict; Invalid channel combination. Valid
configurations are ‘PSCH SSCH P-CCPCH CPICH’ or any
combination of ‘DPCH OCNS and PICH’. The two groups should not
be mixed.
The signal generator is not capable of generating certain channel
combinations. Change the present combination to one of the allowable
combinations suggested by the error message.