54 ErrorMessages
-199 to -100: Command Errors ESG Family Signal Generators
Syntax error
An unrecognized command or data type was encountered. For example, a
string was received when the device does not accept strings.
-101 Invalid character
A syntactic command contains a character that is invalid for that type. For
example, a header containing an ampersand, SETUP&. This error might
be used in place of error numbers -114, -121, -141 and some others.
-100 Command error; Unable to set new frequency with UP/DN command.
The instrument is unable to obtain the necessary information to calculate
the new frequency in compatibility mode with the UP/DN command.
Command error; Unable to set new amplitude with UP/DN command.
The instrument is unable to obtain the necessary information to calculate
the new amplitude in compatibility mode with the UP/DN command.
Command error; Remote active function IS not available.
The instrument only supports the Increment Set (IS) command for
Frequency (FR) and Amplitude (AP) in compatibility mode. All other uses
of IS are invalid.
Command error; R0:No standby mode allowed.
In 8656/57 compatibility mode, R0 was received via GPIB. This command
is not supported by the compatibility mode.
Command error; Remote active function DN/UP not available.
In 8656/57 compatibility mode, either DN or UP was received via GPIB.
These commands are not supported by the compatibility mode.
Command error; LO: No low bandwidth ALC mode allowed.
In 8656/57 compatibility mode, LO was received via GPIB. This command
is not supported by the compatibility mode.
Command error
This is a generic syntax error for devices that cannot detect more specific
errors. The code indicates only that a command error as defined in
IEEE 488.2, 11.5.1.1.4 has occurred.