Agilent Technologies E4400-90333 Portable Generator User Manual


 
38 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
Settings conflict; Frequency list and power list are of
unequal size. Turn one list off, set one to size one, or make
their sizes equal.
The frequency list has more than one element and the power list has more
than one element, and they are not of equal size. If any of the frequency,
power, or dwell lists have more than one element, they must all have the
same number of elements. A list of a single element is the same as a list of
an equal size with the single element repeated the necessary number of
times.
Settings conflict; The frequency spacing has been limited by
the number of tones.
The product of the multitone number of frequencies and frequency spacing
must be less than or equal to 15 MHz.
Settings conflict; FSK frequency deviation cannot exceed the
symbol rate.
If the FSK frequency deviation exceeds the symbol rate, the deviation is
changed to match the value of the symbol rate.
Settings conflict; GSM BERT requires Multiframe PN Sequence in
Timeslot E field.
An unsupported data type was encountered. The measurement timeslot
requires a multiframe PN sequence in the E field.
Settings conflict; GSM BERT requires Normal as Timeslot Type.
An unsupported timeslot type was encountered. The measurement
timeslot must be designated Normal.
Settings conflict; High Amplitude is less than Low Amplitude.
The specified High Amplitude value is smaller than the Low amplitude
value for a Sensitivity Search MEASurement? query. No measurement
is performed.
Settings conflict; In this configuration the parameter is a
constant. It may not be set.
An attempt was made to enter a value into a fixed parameter for the
current radio configuration or channel setup. For example, the frame
length parameter for the R-FCH has a fixed value in RC1 and RC2.
However, in RC3 and RC4 this parameter is adjustable.