Agilent Technologies E4400-90333 Portable Generator User Manual


 
Error Messages 23
ESG Family Signal Generators -299 to -200: Execution Errors
-252 Missing media
A legal program command or query could not be executed because of
missing media, for instance no disk in the disk drive. The definition of
what constitutes missing media is device-specific.
Missing media
If this occurs during a memory catalog display, it means the default
memory system could not be located. The instrument is likely not
functioning properly. Report this error to the nearest Agilent Technologies
sales or service office.
-250 Mass storage error
A mass storage error has occurred. This message is used when a device
cannot detect the more specific errors described for errors -251 through
-259.
Mass storage error; EEPROM write timeout on <filename>.
The system was not able to program new data to an EEPROM. The system
is still functional, but files written to EEPROM (such as updated
calibration data) may be lost when the instrument’s line power is cycled.
Report this error to the nearest Agilent Technologies sales or service office.
-241 Hardware missing
A legal program command or query could not be executed because of
missing device hardware. For example, an option was not installed.
Hardware missing; <card_name>
A test communication to a hardware card failed. The instrument is most
likely not functional. Report this error to the nearest Agilent Technologies
sales or service office.
Hardware missing; Installed option boards do not match
configuration information.
A set of option boards have been installed that do not match the
information that was given to the instrument as part of the installation. If
this is the result of a customer installed option, the wrong option was
specified during installation. If this is seen at any other time, the likely
cause is an EEPROM failure on the option card.