44 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
Settings conflict; The selected external trigger setting
conflicts with the previous setting.
The external trigger has been set to positive edge for one trigger source
and negative edge for another trigger source.
Settings conflict; Uplink & Downlink setting conflict. All
selected Uplink timeslots are now set to Downlink.
An Uplink timeslot type has been selected while another timeslot(s) has
Downlink selected. In NADC and PDC digital modulation for bursted data,
only either Uplink (base to mobile) or Downlink (mobile to base) Traffic
Channel type can be selected for all timeslots at any one moment.To
continue data transmission, all Uplink timeslots have been changed to the
Downlink configuration.
Settings conflict; The symbol rate is too high for a > 16
symbol filter. Baseband cannot function.
The filter is in a state that is inappropriate for the hardware. Although no
damage to the hardware will occur, the output is in an unknown state.
Report this error to the nearest Agilent Technologies sales or service office.
Settings conflict; Unsupported configuration for Pilot Gating.
Pilot Gating can only be used for the Pilot alone or in
conjunction with REACH or RCCCH.
Changing the R-PICH gating from full is only allowed in three channel
structures: R-PICH, R-PICH with the R-CCCH, or R-PICH with the
R-EACH.
Settings conflict; User FSK file has deviations which are
incompatible with the current symbol rate. The modulation type
has been defaulted.
A frequency in the selected user FSK file exceeds the maximum possible
deviation for the current symbol rate. The default modulation type for the
mode has been selected to resolve the conflict.
Settings conflict; Walsh conflict between: <CHANNEL NAME1> and
<CHANNEL NAME2>.
There is a Walsh collision between two channels specified. Either one
Walsh code is equal to the other, or it falls under the other Walsh's tree. A
Walsh collision results in an invalid signal. Select a different Walsh code
according to Walsh selection rule.