24 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
Hardware missing; no such SIO address
The expected board is not installed. Report this error to the nearest
Agilent Technologies sales or service office.
-240 Hardware error
A legal program command or query could not be executed because of a
hardware problem in the device. The definition of what constitutes a
hardware problem is completely device-specific. This error is used when
the device cannot detect the more specific errors described for errors -241
through -249.
-233 Invalid version
A legal program data element was parsed but could not be executed
because the version of the data is incorrect to the device. This particular
error is used when file or block data elements are recognized by the
instrument, but cannot be executed for reasons of version incompatibility.
For example, a non-supported file version or a non-supported instrument
version.
-232 Invalid format
A legal program data element was parsed but could not be executed
because the data format or structure is inappropriate. For example, when
loading memory tables or when sending a SYSTem:SET parameter for an
unknown instrument.
-231 Data questionable
The measurement accuracy is questionable.
Data questionable; EEPROM copy of <filename>.
The EEPROM copy of a file has a correctable error. The system
automatically performs the correction. A potential cause is a failing
EEPROM. Report this error to the nearest Agilent Technologies sales or
service office.
Data questionable; RAM copy of <filename>.
The non-volatile RAM copy of a file has a correctable error. The system
automatically performs the correction. A potential cause is a failing
backup battery.
-230 Data corrupt or stale
Possibly invalid data. A new reading was started but not completed since
last access.