ESG Family Signal Generators -299 to -200: Execution Errors
Settings conflict; Continuous & Discontinuous setting
conflict. All selected Continuous timeslots are now switched
to Discontinuous.
Indicates that while in TETRA mode, an attempt has been made to
simultaneously assign continuous and discontinuous protocols to the
timeslots. Continuous and discontinuous protocols cannot coexist in
TETRA mode. If a discontinuous protocol is selected for any timeslot, all
timeslots’ protocols are changed to discontinuous protocols.
Settings conflict; Current frame length does not have symbol
boundary. End of frame has been patched with extra bits to
allow an integer of symbols in a frame.
With Option UN8, in TDMA format you can select from many different
modulations with different bits per symbol. Because all TDMA formats
have a fixed frame size, in some modulations end-of-frame and
end-of-symbol do not happen at the same time. This message indicates
that end-of-frame has been patched with extra bits to allow an integer of
symbols per frame.
Settings conflict; Discontinuous & Continuous setting
conflict. All selected Discontinuous timeslots are now
switched to Continuous.
Indicates that while in TETRA mode, an attempt has been made to
simultaneously assign discontinuous and continuous protocols to the
timeslots. Discontinuous and continuous protocols cannot coexist in
TETRA mode. If a continuous protocol is selected for any timeslot, all
timeslots’ protocols are changed to continuous protocols.
Settings conflict; Downlink & Uplink setting conflict. All
selected Downlink timeslots are now set to Uplink.
A Downlink timeslot type has been selected while another timeslot(s) has
Uplink selected. In NADC and PDC digital modulation for bursted data,
only either Uplink (base to mobile) or Downlink (mobile to base) Traffic
Channel type can be selected for all timeslots at any one moment.To
continue data transmission, all Downlink timeslots have been changed to
the Uplink configuration.
Settings conflict; EDGE BERT requires Multiframe PN Sequence
in Timeslot E field.
An unsupported data type was encountered. The measurement timeslot
requires a multiframe PN sequence in the E field.