16 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
-299 to -200: Execution Errors
An error has been detected by the instrument’s execution control block. Errors in this class
set the execution error bit (bit 4) in the event status register (IEEE 488.2, section 11.5.1).
In this case:
Either a <PROGRAM DATA> element following a header was evaluated by the device
as outside of its legal input range or is otherwise inconsistent with the device’s
capabilities, or
a valid program message could not be properly executed due to some device condition.
Execution errors are reported after rounding and expression evaluation operations are
completed. Rounding a numeric data element, for example, is not reported as an execution
error.
-294 Incompatible type
The type or structure of a memory item is inadequate.
-293 Referenced name already exists
A downloaded program attempted to define an element (a variable,
constant, filename, etc.) that had already been defined.
-292 Referenced name does not exist
A downloaded program attempted to access an undefined element
(a variable, constant, filename, etc.).
-291 Out of memory
A downloaded program required more memory than was available in the
instrument.
-290 Memory use error
A user request has directly or indirectly caused an error related to memory
or <data_handles>. This is not the same as “bad” memory.
-286 Program runtime error
A runtime error was detected in a downloaded program.