ESG Family Signal Generators -399 to -300: Device-Specific Errors
System error; Cannot change manual point until list mode error
condition cleared.
An error is keeping the sweep/list from being able to set the frequency
and/or power. Until the problem is addressed, the manual point cannot be
changed.
System error; File Not Found.
The specified file cannot be found. Verify that the file exists in the
instrument file system.
System Error; FIR subsystem generated more coefficients than
the hardware can handle.
The maximum number of coefficients for a given set of hardware has been
exceeded. The filter must be created with fewer coefficients.
System error; Hardware error. Unable to apply Phase inversion!
There is an error with the UN8 option. Try turning CDMA2000 mode off
and on again to reset the card. If the problem persists, report this error to
the nearest Agilent Technologies sales or service office.
System error; RS232 buffer overflow: character lost.
The RS232 buffer has been exceeded. The most recent character has been
dropped.
System error; Unable to determine which attenuator is
installed.
An invalid attenuator identification code has been detected. Possible
causes include a loose attenuator control cable. The instrument will likely
not produce the proper output power levels. Report this error to the
nearest Agilent Technologies sales or service office.
-300 Device-specific error
This is a generic device-dependent error for devices that cannot detect
more specific errors. The code indicates only that a device-dependent error
as defined in IEEE 488.2, 11.5.1.1.6 has occurred.