Agilent Technologies E4400-90333 Portable Generator User Manual


 
10 ErrorMessages
-399 to -300: Device-Specific Errors ESG Family Signal Generators
Out of memory; Histogram display cannot function.
There is insufficient working memory available. Report this error to the
nearest Agilent Technologies sales or service office.
Out of memory; Impulse response display cannot function.
There is insufficient working memory available. Report this error to the
nearest Agilent Technologies sales or service office.
Out of memory; There is insufficient memory to copy the
current pattern. The secondary pattern is unchanged.
There was insufficient room to copy or create the current data generation
pattern data in order to make it a secondary pattern. To avoid this, set up
the smaller of the two patterns first.
Out of memory; There is insufficient memory to store the
secondary frame in Pattern RAM. The secondary frame has been
turned off.
There was insufficient room to copy or create the current data generation
pattern data in order to make it a secondary pattern. To avoid this, set up
the smaller of the two patterns first.
Out of memory; Insufficient RAM
A memory comparison between a shadow RAM data area and the
corresponding EEPROM data area could not be performed due to
insufficient working RAM. This does not necessarily imply a memory
problem, since this comparison is only used in stringent diagnostic
situations.
Out of memory; I/Q map display cannot function.
There is insufficient working memory available. Report this error to the
nearest Agilent Technologies sales or service office.
Out of memory; List formation
The device was unable to allocate space for a lookup table, such as for list
mode precalculation. List mode cannot run until either fewer frequencies
have been supplied or more memory becomes available and the same set of
frequencies are sent again, FREQ:MODE CW is executed, or :FREQ:MODE
LIST is executed.