26 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
Data corrupt or stale; License file is obsolete, missing
entries defaulted.
This error can only occur during the instrument’s first power up after
having downloaded new firmware. It indicates that the software license
file is missing entries for options supported by the current revision of
firmware. The missing entries have been added to the file, and their
license key values have been defaulted to zero. This error will occurs
during the first power up after downloading new firmware into the
instrument.
Data corrupt or stale; RAM copy of <filename>.
The non-volatile RAM copy of a file is either corrupt or is out of date with
the EEPROM master copy (if one exists). The system automatically
re-initializes the file from EEPROM (if appropriate) or from a default
algorithm. A potential cause is a failing backup battery.
-226 Lists not same length
Attempted to use LIST structure having individual LISTs of unequal
length.
-225 Out of memory
The device has insufficient memory to perform the requested operation.
-224 Illegal parameter value
Used where exact value, from a list of possibilities, was expected.
Illegal parameter value; Command requires at lease one data
value.
A modulation type cannot be created without data.
Illegal parameter value; Total user file bits must be greater
than 0.
An attempt was made to select a user file that contains 0 bits. Select
another user file that contains at least 1 bit.
Illegal parameter value; Total user file bits must be greater
than 0 and smaller than 500000.
The current hardware supports user file size between 1 and 500000 bits.
Select a different file with the correct size.