Agilent Technologies E4400-90333 Portable Generator User Manual


 
Error Messages 55
ESG Family Signal Generators 201 to 702: Device-Specific Errors
201 to 702: Device-Specific Errors
Some device operations did not properly complete, possibly due to an abnormal hardware
or firmware condition. These codes are also used for self-test response errors. Errors in this
class set the device-specific error bit (bit 3) in the event status register (IEEE 488.2,
section 11.5.1).
The <error_message> string for a positive error is not defined by SCPI. A positive error
indicates that the instrument detected an error within the GPIB system, within the
instrument’s firmware or hardware, during the transfer of block data, or during
calibration.
201 Bad file number; Unable to check Data Generator memory.
The instrument was not able to generate the pattern necessary to perform
the data generator memory test. This does NOT imply a data generator
memory failure. Report this error to the nearest Agilent Technologies sales
or service office.
208 I/O Error; <file name>
Invalid file name.
I/O error; Delete a non-saved state register. Delete register
ignored.
The user has attempted to delete an unused (empty) state. This error
message is informational only.
I/O error; Delete empty sequence <sequence_name>. Delete
sequence ignored.
The user has attempted to delete a sequence that is empty. This error
message is informational only. Typically, this error is reported several
times when the “Delete All Sequences” command is executed. If the file is
displayed by Catalog, delete explicitly.
I/O error; Directory does not have a default file type.
Report this error to the nearest Agilent Technologies sales or service office.
I/O error; Directory refuses to release ownership.
Report this error to the nearest Agilent Technologies sales or service office.
I/O error; Save a state register ignored.