Agilent Technologies E4400-90333 Portable Generator User Manual


 
46 ErrorMessages
-299 to -200: Execution Errors ESG Family Signal Generators
-220 Parameter error
A program data element related error has occurred. This particular error
message is used if the device cannot detect a more specific error described
for errors -221 through -229.
Parameter error; <modulation file name>
The specified file is not a modulation file type.
-215 Arm deadlock
The arm source for the initiation of a measurement is set to GET and a
subsequent measurement query is received. The measurement cannot
begin until a GET is received, but the GET would cause an INTERRUPTED
error.
-214 Trigger deadlock
A trigger source for the initiation of a measurement is set to GET and a
subsequent measurement query is received. The measurement cannot
begin until a GET is received, but the GET would cause an INTERRUPTED
error.
Trigger deadlock
TRIGger:SOURce BUS was selected when a MEASurement? query was
executed. No data is returned.
-213 Init ignored
A request for a measurement initiation was ignored as another
measurement was already in progress.
Init ignored; Cannot initiate sweep in manual mode.
The manual mode is on and therefore the instrument cannot sweep.
Init ignored; Sweep is already initiated.
The list/sweep is currently initiated and sweeping, therefore the command
is not legal according to SCPI.
Init ignored; Sweep is already continuously initiated.
The list/sweep is continuously initiated and sweeping, therefore the
command is not legal according to SCPI.