HP (Hewlett-Packard) HP 16555D Weather Radio User Manual


 
Error Messages
Must have at least 1 edge specified.
You must assign at least one clock
edge to one of the available clocks in the Master clocking arrangement. In
addition, if the Slave clock is being used, it must have at least one clock edge
assigned. The analyzer will not let you close the clock assignment pop-up
until an edge is specified.
Time correlation of data is not possible.
Before time correlation of data
is possible, time tags must be placed on the data. If you want time correlated
data, set the Count field in the Trigger menu to Time.
Maximum of 32 channels per label.
This message appears when you try
to assign more than 32 channels to a single label
NO DISK.
There are no disks in the disk drives. You must place a disk into
the disk drive before configurations can be loaded or stored.
Timer is off in sequence level.
At least one sequence level has specified a
timer as part of the sequence instruction and that timer is not turned on.
The timer must be set to either Start, Pause, or Continue.
Timer is specified in sequence, but never started.
A timer is specified
somewhere in the sequence, but was not started. The timer must be set to
Start. The timer can be set to Start in any sequence level.
Problems reading file.
The user is trying to translate a configuration file
that can not be opened a second time.
Inverse assembler not loaded - bad object code.
Corrupt inverse
assembler file. Try getting another copy of the inverse assembler file and
loading that.
Insufficient memory to load IAL - load aborted.
There is not a block of
free memory large enough to load inverse assembler.
ASCII entry not available.
The ASCII base is not available. You must use
another base selection.