Time Markers
Time markers are indicators located in the waveform display that are
used as reference marks to obtain time values between markers, or
between each marker and the trigger point.
In a state analyzer, Time markers only become available when the
Count field is set to Time in the Trigger menu.
The markers are color-coded for easy recognition in the waveform
display. The X marker is represented by a vertical green line and the
O marker is represented by a vertical yellow line. In addition, both
markers are labeled at the bottom of the display.
If Pattern markers are assigned, the Time markers are initially placed
at the same locations in the data listing.
Time markers have two fields for control and display:
• Trig to X/Trig to O Fields
• Marker Label/Base and Display
7–17