Seconds Per Division Field (Timing only)
When you set the analyzer Type field in the Configuration menu to Timing,
the analyzer uses its own internal clock. In this mode, the X-axis of the
waveform display is measured in seconds per division (sec/Div).
Use the sec/Div field to select the seconds per division resolution of the
X-axis. The range of the sec/Div field is 1 ns/Div to 8.8 s/Div for the
HP 16554A, 4.4 s/Div for the HP 16555A, or 2.2 s/Div for the HP 16555D. In
half-channel mode, the maximum sec/Div setting is reduced by half. You set
the sec/Div field either by touching the sec/Div field and rotating the knob, or
by touching the sec/Div field twice and then using the pop-up keypad.
When using the knob to set the sec/Div, the value will change in a 1-2-5
sequence. By adjusting the sec/Div, you can zoom in to view a desired part of
the display.
Seconds Per Division Field
Seconds per Division
field
The Waveform Menu
Seconds Per Division Field (Timing only)