Emerson 2468CB Weather Radio User Manual


 
Hydrastep 2468CB & 2468CD Manual Remote Display Options 24683B, C & D
24685034 4-5
4.2.2 RECONFIGURING THE 24683D REMOTE DISPLAY
The procedure for reconfiguring the 24683D for operation with 18 to 32 electrodes is:
1. Undo the four screws securing the transparent cover and remove the cover.
2. Undo the four screws securing the front panel and lay the panel carefully away from the
top of the box, still connected by the ribbon cable.
3. Carefully remove the link headers from sockets LK2 and LK4 on the decoding PCB, by
easing them out with a suitable flat-bladed screwdriver. (See Figure 4.1)
4. Fit the link headers, removed in step 3, into sockets LK1 and LK3.
5. Refit the front panel and the transparent cover.