Agilent Technologies 16750A Weather Radio User Manual


 
85
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
Saving/Recalling Trigger Setups
You can save a trigger setup within a session by using trigger save/
recall.
“To save a trigger setup” on page 85
“To recall a trigger setup” on page 85
“To clear the trigger save/recall list” on page 85
See Also “Save/Recall Subtab” on page 157
To save a trigger setup
1. Set up the trigger.
2. In the Trigger tab's Save/Recall subtab, select the Save button.
3. Select a memory location to store the trigger setup in.
4. In the Buffer Name dialog, enter a descriptive name for the trigger setup.
To recall a trigger setup
1. In the Trigger tab's Save/Recall subtab, select the Recall button.
2. Choose the trigger setup from one of the previous measurements or one of
the save/recall memories.
Recalling a trigger setup changes the trigger arming, memory depth,
and trigger position as well as the trigger sequence. Recalling a trigger
setup will not change the sampling mode configuration.
If one of the settings in the recalled trigger setup conflicts with the
sampling mode configuration, it will be set to the closest setting.
Also, if the trigger setup uses a trigger function library that does not
exist on this mainframe, it will not load correctly.
To clear the trigger save/recall list
1. When the Trigger tab is displayed, select Save/Recall Memories from the
Clear menu.