An eye finder measurement is currently running. Stop the eye finder
or wait for it to complete before running the eye finder.
The eye finder is already running on the other machine defined for
this analyzer. Eye finder cannot run on both machines at the same
time.
"Cannot run the Logic Analyzer at this time."
An eye finder measurement is currently running. Stop the eye finder
measurement or wait for it to complete before running the logic
analyzer measurement.
Eye finder uses the same hardware as an ordinary logic analyzer
measurement uses. Therefore, they cannot be performed at the same
time.
"Characterizer cannot be loaded"
This is an internal error. It means that a software unit responsible for
measuring one of the channels did not fit the hardware setup and the
other characterizers which were already loaded. Try running eye
finder on just one label (or one channel) at a time to attempt to clear it
up. Contact support if this persists.
"Complete: DATE"
The measurement completed successfully on the date and time given.
This time may also be accessed by expanding a label in the results
display and selecting the clock icon or selecting "Show Time Stamp" on
the popup menu raised by selecting the channel name in the results
display.
"Eye Finder only operates when the analyzer clocking is master clock (slave and/or
demux clocking are not supported)."
See the "Clock Setup" section of the "Sampling" tab in the analyzer
setup window.
"Eye Finder only operates when the analyzer is setup for state analysis."
See the "Pod Assignment" dialog accessed from the "Format" tab in the
analyzer setup window.