Fluke 2635A Lawn Mower User Manual


 
Appendices
Specifications
A
A-11
Resistance Measurements
Resolution
See Table A-13.
Table A-13. Resistance Measurements - Resolution.
Range Resolution Typical Full
Scale Voltage
Maximum
Current
Through
Unknown
Maximum
Open Circuit
Voltage
Slow Fast
300 10 m 0.1 0.22V 1 mA 3.2V
3 k 0.1 1 0.25V 110 µA 1.5V
30 k 1 10 0.29V 13 µA 1.5V
300 k 10 100 0.68V 3.2 µA 3.2V
3 M 100 1 k 2.25V 3.2 µA 3.2V
10 M 1 k 10 k 2.72V 3.2 µA 3.2V
Accuracy
See Table A-14.
Table A-14. Resistance Measurements - Accuracy (Four-Wire)
4-Wire Accuracy ±(%±)
Range
18°C to 28°C
0°C to 60°C
90 Days, Slow 1 Year, Slow 1 Year, Fast 1 Year, Slow 1 Year, Fast
300 0.013% + 20 m 0.014% + 20 m 0.014% + 0.2 0.031% + 20 m 0.031% + 0.2
3 k 0.015% + 0.2 0.016% + 0.2 0.016% + 2 0.039% + 0.2 0.039% + 2
30 k 0.013% + 2 0.014% + 2 0.014% + 20 0.039% + 2 0.039% + 20
300 k 0.020% + 20 0.021% + 20 0.021% + 200 0.050% + 20 0.050% + 200
3 M 0.059% + 200 0.063% + 200 0.063% + 2 k 0.231% + 200 0.231% + 2 k
10 M 0.168% + 2 k 0.169% + 2 k 0.709% + 20 k 0.573% + 2 k 0.923% + 20 k