Agilent Technologies 45A Sprinkler User Manual


 
Chapter 4: Calibrating and Adjusting
To check the flat panel display (FPD)
4–7
To check the flat panel display (FPD)
No equipment is required for this procedure. Specifications for flat-panel displays used in the
Infiniium oscilloscope are shown in the following table.
Flat-Panel Display Specifications
1
Enable the graphical interface.
2 Select Self Test from the Utilities menu.
3 Click the Service Extensions box in the Self Test dialog.
4 Select Screen from the Test drop-down list box.
See figure 4-2.
Defect Type Limit
Polarizer Scratch Width 0.05 mm
Length 10 mm
Dent φ ≤ 0.4mm
Dot Defect
(A dot is defined as 1, 2, or 3 stuck
subpixels touching horizontally.
Subpixels are horizontal red, green,
blue triads, so these may show up as
one of 8 colors or black.)
Bright dot N 5
Dark dot N 7
Total dot N 12
Two adjacent dots
Bright dot
Dark dot
2 pairs
2 pairs
Three or more adjacent dots Not allowed
Distance between defects
Bright dot
Dark dot
10 mm
10 mm
Line Defect Not allowed
Non-uniformity Check other specifications
Luminance
(The measurement is perpendicular
to the screen surface in both axes.)
Minimum
Typical
160 cd/m
2
200 cd/m
2
When to Use this Procedure
This procedure should not be performed as a part of routine maintenance. Perform the procedure only