GE PT878GC Portable Generator User Manual


 
April 2004
9-44 Servicing the PT878GC
Updating Software Via IrOBEX
To begin updating software via IrOBEX, scroll down to the Flash
Update entry in the Service Menu and press
[ENTER]. The screen
appears similar to Figure 9-27 below.
Figure 9-27: The Flash Update Window
Press [F2] (Cancel) to return to Operate Mode without
erasing the program.
Press [F3] (OK) to confirm that you wish to erase the
program.
4. The meter asks for confirmation. Repeat the options shown in
Step 1 above. After the PT878GC reboots, the screen appears
similar to Figure 9-28 below.
Figure 9-28: Flash Reloading Window
Update Flash
OK
--WARNING--
This will erase the
instrumentation coding.
Choose ‘Cancel’ to exit or ‘OK’
to continue.
Cancel
GE Panametrics PCI Loader v3.0 2/26/04
[HW Rev3+]
Reload Flash via Infrared (Yes/No)?
Backup Battery: OK
Firmware CRC: Valid
IrOBEX No IrCOMM