Calibrating the Thickness Gauge Transducer
To zero the transducer offsets, go to the Zero tab. Press the [W ]
or [
X] arrow key until you reach the Zero tab and press [ENTER].
The screen appears similar to Figure 9-9 below.
Figure 9-9: The Zero Window
1. Press the [
T] arrow key to enter the window.
2. The first prompt asks you to choose between single and dual-
point calibration. Use the [
W ] and [X] arrow keys to move
to the appropriate radio button and press
[ENTER].
Note: GE Panametrics recommends dual calibration. Dual-
point calibration is more reliable, but it requires two
known thicknesses of the calibration material.Use single-
point calibration only if you know the sound speed in the
calibration block material to a high degree of accuracy.
See Appendix C for more information.
3. The next prompt asks you to enter the length of Block 1. (If
you are using a GE Panametrics-supplied test block, the length
is printed on the block.) Press
[ENTER] to open the text box.
Display
Thickness Gauge Measure
Cancel OK
Graph
Velocity
Zero
Material
Block 1 Length
Calculated Current
Transducer Delay,
0
0
Single
Dual Calibration
0
mm
Calibrate
µs