GE PT878GC Portable Generator User Manual


 
Servicing the PT878GC 9-39
April 2004
Setting Impulse Response
The Impulse Response option enables you to force the meter to
transmit in one direction only, without changing the AGC setting.
You can then diagnose problems with transducer, pipe, or fluid
configurations. To enter the option, scroll down to the Impulse
Response entry in the Service Menu and press
[ENTER]. The
screen appears similar to Figure 9-24 below.
Figure 9-24: The Impulse Response Option
1. The first prompt asks in which direction you wish the meter to
transmit, upstream or downstream.
a. Press
[ENTER] to open the drop-down list.
b. Use the [
T] or [S] arrow keys to scroll to the desired
direction.
c. Press
[ENTER] to confirm your selection.
Transmit
Stop Exit
Impulse Response
.
Direction Upstream
Percent Gain
20 %
Stopped.