Anritsu 683XXB Portable Generator User Manual


 
A-192 682XXB/683XXB MM
TEST MODEL
RECORD 68277B/68377B
3-8 Spurious Signals Test: RF Output Signals <2 GHz
Test Procedure
Measured Value Upper Limit
Set F1 to 10 MHz
Record the presence of the worst case harmonic
..........
Record the presence of the worst case non-harmonic
.......
_______________dBc
_______________dBc
–30 dBc
–40 dBc
Set F1 to 20 MHz
Record the presence of the worst case harmonic
..........
Record the presence of the worst case non-harmonic
.......
_______________dBc
_______________dBc
–30 dBc
–40 dBc
Set F1 to 30 MHz
Record the presence of the worst case harmonic
..........
Record the presence of the worst case non-harmonic
.......
_______________dBc
_______________dBc
–30 dBc
–40 dBc
Set F1 to 40 MHz
Record the presence of the worst case harmonic ..........
Record the presence of the worst case non-harmonic .......
_______________dBc
_______________dBc
–30 dBc
–40 dBc
Set F1 to 350 MHz
Record the presence of the worst case harmonic ..........
Record the presence of the worst case non-harmonic .......
_______________dBc
_______________dBc
–40 dBc
–40 dBc
Set F1 to 1.6 GHz
Record the presence of the worst case non-harmonic ....... _______________dBc –40 dBc
Set F1 to 1.6 GHz
Record the level of the harmonics of the 1.6 GHz carrier:
3.2 GHz (2nd harmonic)
......................
4.8 GHz (3rd harmonic)
......................
_______________dBc
_______________dBc
–40 dBc
–40 dBc