682XXB/683XXB MM A-45
3-11 Power Level Accuracy and Flatness Tests (Continued)
(Model 68247B/68347B with Option 15B High Power & without Option 2A Step Attenuator)
Power Level Accuracy Test Procedure
Set F1 to 1.0 GHz Set F1 to 5.0 GHz
Set L1 to: Measured Power * Set L1 to: Measured Power *
+13 dBm
+12 dBm
+11 dBm
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7 dBm
+ 6 dBm
+ 5 dBm
+ 4 dBm
+ 3 dBm
+ 2 dBm
+ 1 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
+17 dBm
+16 dBm
+15 dBm
+14 dBm
+13 dBm
+ 12dBm
+11 dBm
+10 dBm
+ 9 dBm
+ 8 dBm
+ 7dBm
+ 6 dBm
+ 5 dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
_________dBm
* Specification is 1.0 dB. * Specification is 1.0 dB.
Power Level Flatness Test Procedure (Step Sweep)
Set L1 to: Max Power Min Power Variation **
+13 dBm _________dBm __________dBm _________dB
** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 1.6 dB (0.05 to 20 GHz).
Power Level Flatness Test Procedure (Analog Sweep) (Model 68347B only)
Set L1 to: Max Power Min Power Variation ***
+13 dBm _________dBm __________dBm _________dB
*** Maximum variation is 4.0 dB (0.01 to 0.05 GHz); 2.0 dB (0.05 to 20 GHz)(typical, not a specification).
TEST MODEL
RECORD 68247B/68347B