10. In the Properties window, make the following se-
lections:
B
its per second 19200 (See Note)
D
ata bits 8
P
arity None
S
top bits 1
F
low control Xon / Xoff
11. After making the selections, click on the OK but
-
ton.
12. Press <ENTER> on the keyboard.
13.
Verify that the
$ prompt appears on the PC dis
-
play.
14. This completes the initial setup for calibration.
4-12 682XXB/683XXB MM
CALIBRATION INITIAL SETUP
NOTE
When calibrating instruments with
operating firmware below Version
2.00, set the Bits per second rate to
9600. (This does not apply to
682X5B/683X5B models.)