3-20
Function Reference
Current Source Setup
Current Source Setup
Accesses the Current Source Setup panel that allows you to turn on or off the
current source, set the current limit, enable pulse width, duty ratio, and syn-
chronize the ADC sync output. Refer to Operating the Internal White Light
Source on page 2-4
Key Path System > Options > Current Source Setup
Related Functions ADC sync, current limit, pulse width, duty ratio
Remote
Commands
SOURce:CURRent:PULSe:STATe ON|OFF
Date/Time (Display Setup Panel)
Turns the date and time on the display on or off. When on, the date and time
will appear on the displays lower, right corner of the display and on printouts.
Key Path System > More System Functions > Display Setup
Related Functions SYSTem:TIMe
Default Math Trace C
Defines the math expression to be used and turns the math operation on:
C=A
LOG
-B, C=A
LOG
+B, C=A
LIN
-B, or C=A
LIN
+B. The result is placed in Trace C.
The math operation is performed in linear units.
See Using Trace Math to Measure Wavelength Drift on page 2-29.
Key Path Traces > Trace Math Off > Default Math Trace C
Remote
Commands
CALCulate3:MATH:STATE OFF|ON|0|1