Agilent Technologies 86140B Yard Vacuum User Manual


 
5-23
Status Listings
OSA Warnings
5062 desc = ADC Triggered Sweep Too Fast
help = A sweep was taken with one of the ADC trigger modes enabled. The
sweep rate was too fast to allow trace data to be acquired for every
wavelength. To resolve this problem increase sweep time or increase the
frequency the external trigger input signal.
5063 desc = Key disabled during applications
help = The key you pressed is not active while an application is running. Exiting
the application should re-enable the key.
5064 desc = Invalid measurement file
help = An attempt was made to recall a measurement file which is contains
invalid or corrupt data.
5065 desc = Error detected in DSP sub-system
help = The Digital Signal Processor has reported an unexpected error. Please
record the hexadecimal number listed with the error and cycle power. If the error
persists, contact the Agilent support center nearest your location.
5066 desc = Error in Enhanced Wavelength Calibration
help = Enhanced Wavelength Calibration failed. Check OSA system. If the error
persists, contact the Agilent Service Center.
5067 desc = OSNR marker not allowed.
help = A request was made to enable the OSNR marker readout. The readout
cannot be enabled for one of the following reasons: OSNR markers cannot be
used on traces which have X axis units other than meters. This typically occurs
in zero span where the X axis units are seconds. OSNR marks cannot be used on
traces which have Y axis units other than power. this typically happens when
trace math is on and the trace Y axis represents a ratio.
5068 desc = The configured SHARE does not exist.
help = A request was made to a remote file or printer share that does not exist
or cannot be found.
5069 desc = The configured SHARE cannot be accessed.
help = A request was made to a remote file or printer share that cannot be
accessed using the configured USER, PASSWORD, or DOMAIN/WORKGROUP.
5070 desc = Trajectory add failed.
help = The trajectory add procedure failed. This usually occurs when the
trajectory table is full or the computed trajectory table is invalid. The correct this
problem, try AUTO ALIGN PRESET and the AUTO ALIGN & ADD TO TRAJECTORY
procedure again. If the error persists, contact the Agilent Service Center.
Table 5-6. OSA Warnings (7 of 12)
Error Number Error Description [description/explanation/examples]