Agilent Technologies 86140B Yard Vacuum User Manual


 
7-2
Specifications and Regulatory Information
Specifications and Regulatory Information
Specifications and Regulatory Information
This chapter contains specifications and characteristics for Agilent 8614xB
optical spectrum analyzers (OSA).
The specifications in this chapter apply to all functions autocoupled over the
temperature range 0
°C to +55°C and relative humidity < 95% (unless otherwise
noted).
All specifications apply after the instruments temperature has been stabilized
after 1 hour continuous operation and the auto-align routine has been run.
Unless otherwise noted, specifications apply without USER CAL.
All specifications apply to measurements made without using the front-panel
connector saver.
Calibration Cycle This instrument requires periodic verification of performance. The instrument
should have a complete verification of specifications at least once every two
years.