7-2
Specifications and Regulatory Information
Specifications and Regulatory Information
Specifications and Regulatory Information
This chapter contains specifications and characteristics for Agilent 8614xB
optical spectrum analyzers (OSA).
• The specifications in this chapter apply to all functions autocoupled over the
temperature range 0
°C to +55°C and relative humidity < 95% (unless otherwise
noted).
• All specifications apply after the instrument’s temperature has been stabilized
after 1 hour continuous operation and the auto-align routine has been run.
• Unless otherwise noted, specifications apply without USER CAL.
• All specifications apply to measurements made without using the front-panel
connector saver.
Calibration Cycle This instrument requires periodic verification of performance. The instrument
should have a complete verification of specifications at least once every two
years.