Agilent Technologies 86140B Yard Vacuum User Manual


 
5-7
Status Listings
SCPI-Defined Errors
listening formats or whose type is unacceptable to the device.
An unrecognized header was received. Unrecognized headers include incorrect
device-specific headers and incorrect or unimplemented IEEE 488.2 common
commands.
A Group Execute Trigger (GET) was entered into the input buffer inside of an
IEEE 488.2 <PROGRAM MESSAGE>.
Events that generate command errors shall not generate execution errors,
device-specific errors, or query errors; see the other error definitions in this
chapter.
Table 5-1. Command Errors (1 of 4)
Error Number Error Description [description/explanation/examples]
101 desc = Invalid character
help = ““
[A syntactic element contains a character which is invalid for that type; for
example, a header containing an ampersand, SETUP&. This error might be used
in place of errors 114, 121, 141, and perhaps some others.]
102 desc = Syntax error
help = ““
[An unrecognized command or data type was encountered for example, a sting
was received when the instrument does not accept strings.]
103 desc = Invalid separator
help = ““
[The parser was expecting a separator and encountered an illegal character; for
example, the semicolon was omitted after a program message unit, *EMC
1:CH1:VOLTS 5.]
104 desc = Wrong Parameter Type
help = ““
[The parser recognized a data element different than one allowed; for example,
numeric or string data was expected but block data was encountered.]
105 desc = GET not allowed
help = ““
[A Group Execute Trigger was received within a program message (see
IEEE 488.2, 7.7).]