Agilent Technologies 86140B Yard Vacuum User Manual


 
5-25
Status Listings
OSA Warnings
6725 desc = Trace lengths do not match.
help = A math expression could not be evaluated because the inputs have
differing sizes (trace lengths). All inputs to this function must be of the same
size.
6726 desc = Trace lengths do not match.
help = A math expression could not be evaluated because the inputs have
differing sizes (trace lengths). All inputs to this function must be of the same
size.
6727 desc = Y axis counts do not match.
help = A math expression could not be evaluated because the inputs have
differing numbers of Y axis data. All input arguments to this function must have
the same number of Y axis data points.
6728 desc = Incorrect number of inputs for math expression
help = A math expression could not be evaluated because the number of input
arguments to the function is incorrect.
6729 desc = Math expression expects units of dBm.
help = A math expression could not be evaluated because the input argument
does not have the required Y axis units of dBm.
6730 desc = Math expression expects units of watts.
help = A math expression could not be evaluated because the input argument
does not have the required Y axis units of watts.
6731 desc = Illegal combination of trace Y axis units
help = A math expression could not be evaluated because the combination of Y
axis units in the input arguments do not make sense. For example it is valid to
divide watts by watts or to divide watts by a unitless value but it is invalid to
divide a unitless value by watts.
6732 desc = Illegal combination of trace X axis units
help = A math expression could not be evaluated because the input arguments
have differing X axis units. All input arguments must have identical X axis units.
This commonly happens when trace math is attempted between a zero span
trace (time units) and a non-zero span trace (wavelength units).
Table 5-6. OSA Warnings (9 of 12)
Error Number Error Description [description/explanation/examples]