Agilent Technologies 86140B Yard Vacuum User Manual


 
7-11
Specifications and Regulatory Information
Specifications
Table 7-8. General Instrument Specifications
Agilent 86140B, 86141B and 86142B,
86146B Benchtop
Agilent 86143B, 86144B, and 86145B
Portable
Dimensions
222 mm H × 425 mm W × 427 mm D
(8.8 in × 16.8 in × 16.8 in)
163 mm H × 325 mm W × 427 mm D
(6.4 in × 12.8 in × 16.8 in)
Weight
16.5 kg (36 lb) 14.5 kg (31 lb)
Environmental
Temperature
a
a. Floppy disk and printer operating temperature range 0 °C to 45 °C.
Operating: 0°C to 55°C; Storage: 40°C to 70°C
Humidity Operating: < 95% RH; Storage: Non-condensing
Altitude Up to 15,000 feet (4,572 meters)
EMI Conducted and radiated interference is in compliance with
CISPR Publication 11, IEC 801-3, IEC 801-4, and IEC 555-2
Power Requirements
Voltage and Frequency 90 Vac to 260 Vac, 44 to 444 Hz
Maximum Power Consumption 230 W
Computer Interfacing
Remote Control Web enabled controls
Compatibility IEEE-488.1, IEEE-488.2 (100%)
Interfaces GPIB, Parallel Printer Port, External VGA Monitor, Keyboard and Mouse (PS/2)
Floppy Disk 3.5 inch 1.44 MB, MS-DOS®
Data Export Spreadsheet and Word Processor Compatible (CSV ASCII)
Graphics Export CGM, PCL, GIF
Instrument Drivers Universal Instrument Drivers (PNP), Compatible with Agilent VEE,
Labview®, Visual Basic and C
++