Agilent Technologies 86140B Yard Vacuum User Manual


 
5-36
Status Listings
OSA Errors
OSA Errors
Table 5-9. OSA Errors (1 of 2)
Error Number Error Description [description/explanation/examples]
20001 desc = Error detected in ADC sub-system
help = An error has been detected in the Analog-to-Digital converter
subsystem. Please record the hexadecimal number listed with the error and
cycle power. If the error persists, contact the nearest Agilent Technologies
Instrument support center for assistance. In the U.S., call (800) 403-0801. See
the Agilent 86140B Series Users Guide for a listing of the Agilent sales and
service offices.
20002 desc = Error detected in slit positioning system
help = An error has been detected in the motor which controls the resolution
bandwidth slit wheel. Please record the hexadecimal number listed with the
error and cycle power. If the error persists, contact the nearest Agilent
Technologies Instrument support center for assistance. In the U.S., call (800)
403- 0801. See the Agilent 86140B Series Users Guide for a listing of the
Agilent sales and service offices.
20003 desc = Error detected in grating positioning system
help = An error has been detected in the motor which controls the diffraction
grating. Please record the hexadecimal number listed with the error and cycle
power. If the error persists, contact the nearest Agilent Technologies Instrument
support center for assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B Series Users Guide for a listing of the Agilent sales and service
offices.
20004 desc = Monochromator calibration data is invalid
help = Factory calibration data for the monochromator is invalid. Please record
the hexadecimal number listed with the error and cycle power. If the error
persists, contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the Agilent 86140B Series Users
Guide for a listing of the Agilent sales and service offices.