3-91
Function Reference
Switch Path No Auto Align
Switch Path No Auto Align
For 86144B/86146B Filter Mode only.
Selects the 9
µm filter mode path. You should select this function if the instru-
ment has not be moved, subjected to temperature changes >2
° C, turned off
and warmed up for at least an hour at the start of each day or to preserve pre-
vious align data.
Key Path Appl’s > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Related Functions Switch Path Auto Align Now
Remote
Commands
ROUTe:PATH:EXTernal
System
Accesses menus for changing, displaying, selecting, and viewing the instru-
ment’s functions. Some of the functions available are:
• shows HW and critical errors, warnings and notices
• sets a title to the display
• displays the firmware revision
• sets printer setup
• moves the active function area
• displays the instrument state information
• changes the display setup
• changes the time and dates
• selects the power-on state
• performs a factory preset
• upgrades the firmware
• zeros the instrument
• sets the wavelength limit
• sets the TransZ 2-3 lock
• changes the auto measure defaults
• sets the GPIB address
• sets networking parameters