3-25
Function Reference
Exchange Menu
Exchange Menu
Accesses the trace exchange selections which include:
• A Exchange B
• B Exchange C
• C Exchange A
• D Exchange A
• E Exchange A
• F Exchange A
Exchanges the X- and Y-axis data of the two traces. The trace pairs that can be
exchanged are Trace A with any trace, and Trace B with Trace C.
Key Path Traces > Trace Math > Exchange Menu
Remote
Commands
TRACe:EXCHange
Factory Preset (IP)
Performs an instrument preset and clears user settings that are not affected
by Preset. That is, alignment settings, user calibration, multipoint amplitude
correction data. User trace and measurement files will be maintained. See
“Preset” on page 3-65.
Key Path System > More System Functions > Service Menu > Factory Preset (IP)
Related Functions Preset
Remote
Commands
CALibration:ALIGn:PRESet