Agilent Technologies E8247C PSG CW Portable Generator User Manual


 
Chapter 2 45
Basic Operation
Configuring the RF Output
Configuring a Ramp Sweep for a Master/Slave Setup
This procedure shows you how to configure two PSGs and an 8757D to work in a master/slave setup.
1. Set up the equipment as shown in Figure 2-7. Use a 9-pin, D-subminiature, male RS-232 cable with the
pin configuration shown in Figure 2-8 on page 46 to connect the auxiliary interfaces of the two PSGs.
You can also order the cable (part number 8120-8806) from Agilent Technologies.
By connecting the master PSG’s 10 MHz reference standard to the slave PSG’s 10 MHz reference input,
the master’s timebase supplies the frequency reference for both PSGs.
Figure 2-7 Master/Slave Equipment Setup