Agilent Technologies E8247C PSG CW Portable Generator User Manual


 
Chapter 10 199
Troubleshooting
Signal Generator Locks Up
3. Release the Preset key.
4. To continue with the sequence, press
Continue (to abort with no lost files, press Abort).
5. When the sequence concludes:
a. Cycle power.
Cycling power restores all previously installed options. Because calibration files are restored from
EEPROM, you should see several error messages.
b. Perform the DCFM/DCΦM calibration.
Refer to the
DCFM/DCΦM Cal softkey description in the Key Reference.
c. Agilent Technologies is interested in the circumstances that made it necessary for you to initiate this
procedure. Please contact us at the appropriate telephone number listed in Table 10-1 on page 203.
We would like to help you eliminate any repeat occurrences.