Agilent Technologies E8247C PSG CW Portable Generator User Manual


 
Chapter 2 35
Basic Operation
Configuring the RF Output
To Edit List Sweep Points
1. Press
Return > Sweep > Off.
Turning the sweep off allows you to edit the list sweep points without generating errors. If sweep
remains on during editing, errors occur whenever one or two point parameters (frequency, power, and
dwell) are undefined.
2. Press
Configure List Sweep.
This returns you to the sweep list table.
3. Use the arrow keys to highlight the dwell time in row 1.
4. Press
Edit Item.
The dwell time for point 1 becomes the active function.
5. Press
100 > msec.
This enters 100 ms as the new dwell time value for row 1. Note that the next item in the table (in this
case, the frequency value for point 2) becomes highlighted after you press the terminator softkey.
6. Using the arrow keys, highlight the frequency value in row 4.
7. Press
Edit Item > 545 > MHz.
This changes the frequency value in row 4 to 545 MHz.
8. Highlight any column in the point 7 row and press
Insert Row.
This adds a new point between points 7 and 8. A copy of the point 7 row is placed between points 7 and
8, creating a new point 8, and renumbering the successive points.
9. Highlight the frequency item for point 8, then press
Insert Item.
Pressing
Insert Item shifts frequency values down one row, beginning at point 8. Note that the original
frequency values for both points 8 and 9 shift down one row, creating an entry for point 10 that contains
only a frequency value (the power and dwell time items do not shift down).
The frequency for point 8 is still active.
10. Press
590 > MHz.
11. Press
Insert Item > -2.5 > dBm.
This inserts a new power value at point 8 and shifts down the original power values for points 8 and 9 by
one row.
12. Highlight the dwell time for point 9, then press
Insert Item.
A duplicate of the highlighted dwell time is inserted for point 9, shifting the existing value down to
complete the entry for point 10.