92 Chapter 5
Dual Arbitrary Waveform Generator
Arbitrary (ARB) Waveform File Headers
3. Return to the ARB Setup menu: Press Return.
This menu lets you change the current instrument settings. Figure 5-3 shows the ARB Setup softkey
menu and the softkey paths used in steps four through nine.
4. Set the ARB sample clock to 5 MHz: Press
ARB Sample Clock > 5 > MHz.
5. Set the modulator attenuation to 15 dB:
Press
More (1 of 2) > Modulator Atten n.nn dB Manual Auto to Manual > 15 > dB.
6. Set the I/Q modulation filter to a through:
Press
I/Q Mod Filter Manual Auto to Manual > Through.
7. Set marker one to blank the RF output at the set marker point(s):
Press
More (2 of 2) > Marker Utilities > Marker Routing > Pulse/RF Blank > Marker 1.
For information on setting markers, see “Using Waveform Markers” on page 104.
8. Set the polarity of Marker 1 negative: Press
Return > Marker Polarity > Marker 1 Polarity Neg Pos to Neg.
9. Return to the Header Utilities menu: Press
Return > Return > Header Utilities.
Notice that the Current Inst. Settings column now reflects the changes made to the current
signal generator setup in steps 4 through 8, but that the saved header values have not changed (as shown
in Figure 5-4 on page 94).
10. Save the current settings to the file header: Press
Save Setup To Header softkey.
The settings from the Current Inst. Settings column now appear in the Saved Header
Settings column. The file header has been modified and the current instrument settings saved. This is
shown in Figure 5-5 on page 94.
While a modulation format is active, the waveform file (AUTOGEN_WAVEFORM) is playing and you can
modify the header information within the active modulation format. Once you turn the modulation format
off, the header information is available only in the dual ARB player. If you turn the modulation format off
and then on, you overwrite the previous AUTOGEN_WAVEFORM file and its file header. To avoid this,
rename the file before you turn the modulation format back on (see page 103).
When you store the waveform file (see page 103), the header information is stored with the waveform.