Agilent Technologies E8247C PSG CW Portable Generator User Manual


 
28 Chapter 2
Basic Operation
Configuring the RF Output
Configuring the RF Output
This section provides information on how to create continuous wave and swept RF (on page 31) outputs. It
also has information on using a mm-Wave source module to extend the signal generator’s frequency range
(see page 47).
Configuring a Continuous Wave RF Output
These procedures demonstrate how to set the following parameters:
RF output frequency
frequency reference and frequency offset (on page 29)
RF output amplitude (on page 30)
amplitude reference and amplitude offset (page 30)
Setting the RF Output Frequency
Set the RF output frequency to 700 MHz, and increment or decrement the output frequency in 1 MHz steps.
1. Return the signal generator to the factory-defined state: Press
Preset.
NOTE You can change the preset condition of the signal generator to a user-defined state. For these
examples, however, use the factory-defined preset state (set the
Preset Normal User softkey
in the Utility menu to Normal).
2. Observe the FREQUENCY area of the display (in the upper left-hand corner).
The value displayed is the maximum specified frequency of the signal generator.
3. Press
RF On/Off.
The
RF On/Off hardkey must be pressed before the RF signal is available at the
RF OUTPUT connector. The display annunciator changes from RF OFF to RF ON. The maximum
specified frequency should be output at the RF OUTPUT connector (at the signal generator’s minimum
power level).
4. Press
Frequency > 700 > MHz.
The 700 MHz RF frequency should be displayed in the FREQUENCY area of the display and also in the
active entry area.
5. Press
Frequency > Incr Set > 1 > MHz.
This changes the frequency increment value to 1 MHz.