Fluke 105B Telescope User Manual


 
5 - 22
92B/96B/99B/105B
Users Manual
USING THE SCOPE RECORD AND MIN MAX
ENVELOPE FUNCTIONS OF THE
FLUKE 96B, 99B, or 105B
In SCOPE mode, press to access the waveform
recording functions (see Figure 5.16). You can use these
functions to examine waveform changes over time.
Figure 5-16. Recording Menu
The recording functions are as follows:
SCOPE RECORD
This is a fast electronic 'chart recorder' kind of function with
30k of memory. The large memory allows you to record
startup events. The recorder stores 125 successive
screens with single channel use, or 62.5 screens with dual
channel use, within the time base range from 20 ms/DIV to
60 s/DIV.
This gives you a recording time from 24 s to 20 hours with
single channel use, and from 12 s to 10 hours with dual
channel use.
You can use cursors and zoom functions to examine
waveform details.
MIN-MAX ENVELOPE
This function displays changes in the shape of a live
waveform as a grey area around the actual trace. It keeps
track of minimum and maximum values that occur over
time. This can be used for waveforms A and B and A±B.