Fluke 105B Telescope User Manual


 
Using Additional Capabilities 6 - 21
To perform a selective reset, do the following:
1. From the User Options menu, press to open the
MORE OPTIONS menu.
2. Press to open the RESET menu (see Figure 6-12).
3. Choose the reset options which include: the default
settings for Master Reset, or selectively resetting any of
the main modes, e.g., Scope, Meter, Ohm, Diode, or
EXT.mV.
Figure 6-12. Reset Menu Window
4. Press to reset.
Some of the configuration settings can be preset to
become active at every power on.
1. From the User Options menu, press (MORE
OPTIONS), and open the STARTUP MENU window
(see Figure 6-13).
Figure 6-13. Startup Menu Window
2. Specify on or off for each of the following:
- whether or not to identify the waveform (trace)
source (TRACE IDENTIFIERS).
- whether or not to identify the trigger source, slope,
level, and delay (TRIGGER IDENTIFIER)
- whether or not to CLEAR at RUN. When Clear at
Run is set to on, and you perform a single shot, the
old waveform is cleared first each time the test tool
writes a new waveform.
- whether or not to select high or low INFORMATION
LEVEL facility. When the Information Level is set to
high, you will get all on-line information available in
your ScopeMeter test tool.
Factory defaults place all four of these items on.
3. Press to select the new Default Settings.
4. Press to close the window.