Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
Syntax and Commands
2Ć42
*CAL? (Query Only)
Instructs the oscilloscope to perform an internal selfĆcalibration and return its
calibration status.
NOTE
The selfĆcalibration can take three and a half minutes or more to
respond. No other commands will be executed until calibration is
complete.
Calibration and Diagnostic
*CAL?
*CAL ?
<NR1> = 0 indicates that the calibration completed without any errors deĆ
tected.
<NR1> 0 indicates that the calibration did not complete successfully or
completed with errors.
*CAL?
performs an internal selfĆcalibration and might return 0 to indicate that the
calibration was successful.
Group:
Syntax:
Returns: