Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
Syntax and Commands
2Ć178
WFMPre:<wfm>:YMUlt
Sets or queries the vertical scale factor, in YUNit per unscaled data point
value. On input, <wfm> always defaults to the reference location specified
by DATa:DESTination regardless of what is sent.
Waveform
WFMPre:<wfm>:YMUlt <NR3>
WFMPre:<wfm>:YMUlt?
?
<Space> <NR3>
WFMPre :
<wfm>
:
YMUlt
<NR3> is the scale factor, in YUNits (usually volts), per digitizing level.
WFMPre:<wfm>:YOFf
Sets or queries the vertical position of the waveform. On input, <wfm> alĆ
ways defaults to the reference location specified by DATa:DESTination reĆ
gardless of what is sent.
Waveform
WFMPre:<wfm>:YOFf <NR3>
WFMPre:<wfm>:YOFf?
?
<Space> <NR3>
WFMPre :
<wfm>
:
YOFf
<NR3> is the position in digitizing levels.