Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć155
TRIGger:MAIn:HOLdoff:VALue
Sets or queries the main trigger holdoff value.
Trigger
TRIGger:MAIn:HOLdoff:VALue <NR3>
TRIGger:MAIn:HOLdoff:VALue?
<NR3>
TRIGger
:
:
MAIn
HOLdoff
VALue
:
?
<Space>
<NR3> is the holdoff, from 500 ns to 10 s.
TRIGGER:MAIN:HOLDOFF:VALUE 10
sets the holdoff value to 10 s.
TRIGger:MAIn:LEVel
Sets the main trigger level. This command is equivalent to adjusting the
frontĆpanel TRIGGER LEVEL knob.
Trigger
TRIGger:MAIn:LEVel { ECL | TTL | <NR3> }
TRIGger:MAIn:LEVel?
TRIGger :
:
MAIn
LEVel
<Space>
ECL
?
TTL
<NR3>
ECL specifies a preset ECL level of -1.3 V.
TTL specifies a preset TTL level of 1.4 V.
<NR3> specifies the main trigger level, in volts.
TRIGGER:MAIN:LEVEL?
might return 1.4, indicating that the main edge trigger is set to 1.4 V.
Group:
Syntax:
Arguments:
Examples:
Group:
Syntax:
Arguments:
Examples: