Command Descriptions
Syntax and Commands
2Ć74
DIAg:SELect:ALL (No Query Form)
Specifies that all system test sequences will be run when the DIAg:STATE
EXECUte command is sent.
Calibration and Diagnostic
DIAg:SELect:ALL ALL
:
SELect
DIAg : <Space>
ALLALL
ALL selects functional, memory, and register tests for the acquisition, procesĆ
sor and display systems, and self diagnostics for the front panel.
DIAg:STATE (No Query Form)
Executes the diagnostic tests specified by the DIAg:SELect command.
When the test sequence has completed, any of the modules or module
interfaces that failed diagnostics are displayed on the screen and stored in an
internal log file. The pass/fail status is returned by DIAg:RESUlt:FLAg? and
the internal log is returned by DIAg:RESUlt:LOG?. This command is equivaĆ
lent to running Extended Diagnostics by selecting Execute in the Utility menu
when System is set to Diag.
NOTE
The DIAg:STATE EXECute command can take 30 seconds or more to
respond. This command performs a warm boot and does not return
control to the instrument controller until diagnostics are complete.
Calibration and Diagnostic
DIAg:STATE EXECute
:
STATE
DIAg
<Space>
EXECute
EXECute runs the diagnostic test sequences specified by the DIAg:SELect
command. When complete, the oscilloscope returns to the state it was in just
prior to the test. If the PON event was enabled before running the tests, a
Service Request is generated. When the Service Request is received, the
pass/fail status of the tests can be returned by executing DIAg:RESUlt:FLAg?.
The DIAg:STATE EXECute command clears the following:
H the Event Queue
Group:
Syntax:
Arguments:
Group:
Syntax:
Arguments: